►Comparison of Measuring Methods of Corneal Thickness with Orbscan System and Sonogage Pachymeter in Pre and Post LASIK Treatment◄

回雷射近視手術專論─專業論文清單

中華民國眼科學會學術討論會論文摘要

Academic and Annual Meeting of the Ophthalmological Society of Taiwan (R.O.C)

第五屆國際眼科專題演講會暨第四十二次學術演講會(九十年十二月八日~九日)

 ABSTRACT FORM

 Comparison of Measuring  Methods of Corneal Thickness
 with Orbscan System and Sonogage Pachymeter in Pre and Post LASIK Treatment 

張朝凱、陳德照*

台北博仁醫院、長虹診所、德照眼科診所

PurposeTo evaluate corneal thickness with the Orbscan topography system  and the Sonogage ultrasonic pachymeter in pre and post-LASIK eyes to determine the different effects of LASIK on these measurements

MethodWe measure central corneal thickness of right eyes in 30 patients before and after LASIK using both Orbscan and Ultrasonic pachymeter

ResultsThe ultrasound instrument consistently measured a greater corneal  thickness than Orbscan pachymeter .With the retrospectively review , the mean difference between ultrasound and Orbscan pachymeter was 15.6/9.6 and 35.2/18.5 in pre and postoperative eyes respectively ( p 0.092 ).

ConclusionUltrasound pachymeter consistently measures thicker than Orbscan pachymeter ; and there is particular interest as to the reliability of post-LASIK central pachymeter because of concern regarding corneal bed thickness for LASIK enhancements ; therefore , it is reassuring that postoperative Orbscan measurements are consistently lower and more conservative than ultrasound measurements.