►Comparison of Measuring Methods of Corneal Thickness with Orbscan System and Sonogage Pachymeter in Pre and Post LASIK Treatment◄
中華民國眼科學會學術討論會論文摘要
Academic and Annual
Meeting of the Ophthalmological 第五屆國際眼科專題演講會暨第四十二次學術演講會(九十年十二月八日~九日)
ABSTRACT
FORM
Comparison
of Measuring Methods of Corneal Thickness
張朝凱、陳德照*
台北博仁醫院、長虹診所、德照眼科診所*
Purpose:To
evaluate corneal thickness with the Orbscan topography system and the
Sonogage ultrasonic pachymeter in pre and post-LASIK eyes to determine the
different effects of LASIK on these measurements
Method:We
measure central corneal thickness of right eyes in 30 patients before and
after LASIK using both Orbscan and Ultrasonic pachymeter
Results:The
ultrasound instrument consistently measured a greater corneal thickness
than Orbscan pachymeter .With the retrospectively review , the mean
difference between ultrasound and Orbscan pachymeter was 15.6+/-9.6
and 35.2+/-18.5
in pre and postoperative eyes respectively ( p 0.092 ).
Conclusion:Ultrasound pachymeter consistently measures thicker than Orbscan pachymeter ; and there is particular interest as to the reliability of post-LASIK central pachymeter because of concern regarding corneal bed thickness for LASIK enhancements ; therefore , it is reassuring that postoperative Orbscan measurements are consistently lower and more conservative than ultrasound measurements.
|